Model Number MMT-751PNAB
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 03/16/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer's wife reported via phone call the insulin pump had a frozen display and blank display issue.The customer¿s blood glucose was 302 mg/dl at the time of incident and treated with manual injection.Customer declined high blood glucose troubleshooting.Customer stated that the insulin pump was frozen and the buttons were unresponsive.Customer confirmed that the time is not advancing and the buttons remained unresponsive after battery change.Customer had a blank display that was resolved by battery replacement.It was also mentioned that the customer was hospitalized due to high blood glucose.The customer was advised to discontinue use of pump and revert to back-up plan.The insulin pump is being replaced and is expected to return for analysis.
 
Additional Manufacturer Narrative
Insulin pump had blank display due to moisture damage on lcd board.Unable to perform idle current test, run current test, self test, off no power test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test due to blank display.Unable to verify frozen screen, time and button keypad anomaly due to blank display.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751PNAB
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7401597
Report Number3004209178-2018-62424
Device Sequence Number61791
Product Code OZO
UDI-Device Identifier00643169507319
UDI-Public(01)00643169507319
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Patient Family Member or Friend
Type of Report Initial,Followup
Report Date (Section B) 08/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-751PNAB
Device Catalogue NumberMMT-751PNAB
Device Lot NumberA3751PNABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2018
Initial Date Received by Manufacturer 03/16/2018
Supplement Date Received by Manufacturer07/16/2018
Initial Report FDA Received Date04/05/2018
Supplement Report FDA Received Date08/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age65 YR
Patient Weight200
-
-