| Catalog Number |
L880G |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an unspecified plastic surgery procedure the suture was broke as it was being pulled tight through tissue.Procedure was completed with another suture from a different lot.There were no patient consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # : (b)(4).Representative samples were returned for evaluation.They were visually and functionally examined for tensile strength and they met the requirements.Per the condition of the representative samples, no performance - breakage suture was found on the suture and the tested sample met the finished goods requirements.In addition, a review of the batch manufacturing records for the possible batch number was conducted and the batches met all finished goods release criteria.
|
| |
|