Model Number N/A
Health Effect - Clinical Code No Information (3190)
Date of Event 04/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The difference in the readings was determined to be clinically significant.Per label copy/ package insert - high or low blood glucose results can indicate potentially serious medical conditions.In case of an unexpected result, you should repeat the test using a new test strip.If the result is still unexpected, or the reading is not consistent with how you feel, contact your hcp and treat as prescribed.Any change in the treatment of your diabetes should be discussed with your hcp.Nova max test strip insert- quality control - checking the system control solution test: the nova max control solution is used as a quality control check to make sure that your blood glucose monitor and the nova max glucose test strips are working correctly.Do a control solution test: each time you open a new vial of test strips.- storage and handling keep the nova max glucose test strips vial tightly closed when not in use.Test strips should be stored only in the original vial.Meter and test strips are expected to be returned for evaluation.
 
Event or Problem Description
Complainant called in concerned that his 2 nova max plus meters are reading differently.Back up meter (b)(4) with test strips 1020916267 perform as intended.Meter (b)(4) and test strips from the same vial 1020916267.
 
Additional Manufacturer Narrative
The alleged deficiency could not be confirmed.Although the reported results were found in the memory of the returned device.Failure analysis of the returned product revealed that the nova max plus glucose monitoring device performed within specification.The complainant returned blood glucose test strips from the lot in question.Visual as well as chemical evaluation by testing with control solution and blood, was not able to replicate the alleged issue as the performance testing revealed the strip to perform within specification as well.The dhr was complete and contained all relevant data indicating the product put into finished goods met all specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVA MAX PLUS BLOOD GLUCOSE MONITOR
Common Device Name
BLOOD GLUCOSE MONITOR
Manufacturer (Section D)
NOVA BIOMEDICAL CORPORATION
200 prospect street
waltham MA 02454
MDR Report Key7402565
Report Number3004193489-2018-00001
Device Sequence Number19106930
Product Code NBW
UDI-Device Identifier00385480434355
UDI-Public00385480434355
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K091547
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 04/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/30/2018
Device Model NumberN/A
Device Catalogue Number43435
Device Lot Number1020916267
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/17/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/03/2018
Supplement Date Received by Manufacturer04/03/2018
Initial Report FDA Received Date04/06/2018
Supplement Report FDA Received Date04/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-