| Catalog Number |
S3173355 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's picc was "complicated by a fracture of an external component, requiring replacement.".No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a leak is confirmed.One 3 fr sl power picc catheter was returned for investigation.The catheter appeared to have been trimmed 1 cm from the distal end of the molded wing.The catheter was flushed with water using a 12 ml syringe and a leak was observed at the hub/extension tubing interface.Microscopic observation revealed a partial split in the extension tubing.The fracture surface of the split was rough and exhibited cracks/fissures and beach marks, which are indicative of material fatigue.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's picc was "complicated by a fracture of an external component, requiring replacement." no patient injury reported.
|
| |
|