Catalog Number 400SMTXSFT0406
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the pet lock was intact and slightly bent on the proximal end of the pusher assembly, and the pet lock was stuck inside the introducer sheath friction lock.The embolization coil was intact with the pusher assembly.Conclusions: evaluation of the returned device revealed the pusher assembly pet lock was stuck within the introducer sheath friction lock.The friction lock inner diameter (id) is smaller than the rest of the introducer sheath, which allows the introducer sheath to properly seat on the pusher assembly mid-joint.If the pet lock becomes bent, the smaller id of the friction lock may cause increased friction when passing over the pet lock, and may prevent the introducer sheath from being removed.The non-penumbra microcatheter mentioned in the complaint was not returned for evaluation.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using penumbra smart coils (smart coils).During the procedure, the physician successfully advanced the smart coil into the non-penumbra microcatheter, and then attempted to remove the introducer sheath from the proximal end of the pusher assembly.However, the introducer sheath became stuck on the proximal end of the pusher assembly, and was therefore unable to be removed.The physician therefore removed the smart coil and continued the procedure using a new coil and the same microcatheter.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7403735
Report Number3005168196-2018-00720
Device Sequence Number1963413
Product Code HCG
UDI-Device Identifier00814548015385
UDI-Public00814548015385
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date05/23/2022
Device Catalogue Number400SMTXSFT0406
Device Lot NumberF77300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/08/2018
Initial Report FDA Received Date04/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
-
-