• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE TEMPERATURE-SENSING PEDIATRIC FOLEY CATHETER; FOLEY CATHETER (SILICONE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE TEMPERATURE-SENSING PEDIATRIC FOLEY CATHETER; FOLEY CATHETER (SILICONE) Back to Search Results
Model Number 119110
Medical Device Problem Code Mushroomed (2987)
Health Effect - Clinical Codes Pain (1994); No Consequences Or Impact To Patient (2199)
Date of Event 03/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that there was a cuff-roll on the balloon of the catheter.The patient experienced self limited pain.
 
Additional Manufacturer Narrative
The reported issue was confirmed as cause unknown.Per the visual inspection, the balloon mushroom was noted.Per the functional evaluation, the balloon was inflated with air and deflated and a cuff roll was formed.A 10cc mixture of water and blue methylene was injected using a syringe; the catheter was place for 3 minutes on a flat surface where it deflated un it's own and a cuff roll was formed.The dimensional evaluation the catheter was found within specification.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "to deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Visually inspect the product for any imperfections or surface deterioration prior to use".
 
Event or Problem Description
It was reported that there was a cuff-roll on the balloon of the catheter.The patient experienced self limited pain.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® ALL-SILICONE TEMPERATURE-SENSING PEDIATRIC FOLEY CATHETER
Common Device Name
FOLEY CATHETER (SILICONE)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7404508
Report Number1018233-2018-01077
Device Sequence Number16289015
Product Code EZL
UDI-Device Identifier00801741046001
UDI-Public(01)00801741046001
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K070582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,other
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Model Number119110
Device Catalogue Number119110
Device Lot NumberNGBW0780
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2018
Initial Date Received by Manufacturer 03/14/2018
Supplement Date Received by Manufacturer05/17/2018
Initial Report FDA Received Date04/06/2018
Supplement Report FDA Received Date05/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-