• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HALYARD - IRVINE ON-Q C-BLOC WITH DUAL SELECT-A-FLOW 600 ML X 1-7 ML/HR + 1-7 ML/HR; ELASTOMERIC - SAF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HALYARD - IRVINE ON-Q C-BLOC WITH DUAL SELECT-A-FLOW 600 ML X 1-7 ML/HR + 1-7 ML/HR; ELASTOMERIC - SAF Back to Search Results
Model Number CB6007
Medical Device Problem Code Excess Flow or Over-Infusion (1311)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is not possible as no lot number was provided.Root cause could not be determined.All information reasonably known as of 03-apr-2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.(b)(4).
 
Event or Problem Description
Fill volume: unknown, flow rate: 6 ml/hr, procedure: midline laparotomy, cathplace: incision.It was reported that an elastomeric pump was titrated to 6 ml/hr and it emptied in 24 hours.The pump was incisionally placed for a midline laparotomy.The was no reported patient injury.
 
Additional Manufacturer Narrative
The evaluation summary concluded that a fast flow was not observed.The pump was refilled to nominal volume with a baxa repeater pump and infused at a all selectable rates.Flow accuracy testing was performed with the saf set to 6ml/hr.The second saf flow rate was below specification which may be attributed to the pump being filled more than once.Pressure pot on the saf flow rates was performed and all of the tested rates on saf #1 were within specification with a +/-20%.Pressure pot testing on saf #2 met specification except for the 2ml/hr line.A small amount of drug crystallization was found in the 2ml/hr line.The patient was not taking any other medication during use of the pump.The device history record for the reported lot number, 0202848438, was reviewed and documented that the lot was manufactured according to the approved manufacturing procedures, specifications, and then released by quality assurance.Root cause could not be determined.All information reasonably known as of 29-may-2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).
 
Additional Manufacturer Narrative
The actual device was returned for evaluation.All information reasonably known as of 18-apr-2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ON-Q C-BLOC WITH DUAL SELECT-A-FLOW 600 ML X 1-7 ML/HR + 1-7 ML/HR
Common Device Name
ELASTOMERIC - SAF
Manufacturer (Section D)
HALYARD - IRVINE
43 discovery
suite 100
irvine CA 92618
MDR Report Key7404671
Report Number2026095-2018-00044
Device Sequence Number11250094
Product Code MEB
UDI-Device Identifier30680651134777
UDI-Public30680651134777
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K063530
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date05/05/2020
Device Model NumberCB6007
Device Catalogue Number101347700
Device Lot Number0202848438
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/13/2018
Supplement Date Received by Manufacturer04/09/2018
05/08/2018
Initial Report FDA Received Date04/06/2018
Supplement Report FDA Received Date04/20/2018
05/29/2018
Patient Sequence Number1
-
-