| Model Number |
9525 |
| Medical Device Problem Code |
Cut In Material (2454)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
03/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Verified complaint of "exposed copper wire".The power cord was examined and the outer shielding contains a cut mark and the fibers are shredded.
|
| |
|
Additional Manufacturer Narrative
|
|
N evaluation of the scd 700 was performed and the customer states ¿exposed copper wire.¿ the unit was triaged and the customer¿s reported condition was confirmed.The power cord was examined and the outer shielding contains a cut mark and the fibers are shredded.The potential root cause is customer misuse due to running over the cord with hospital bed or the procedure used to unplug the unit.Device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.A review of the device history record shows that this unit was manufactured and was released meeting all manufacturing specifications.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
|
| |
|