Catalog Number 55840005550
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Pain (1994)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient pre-operatively had lumbar back pain and underwent lumbar posterior fixation at l3/4/5.Post-op, the screws at l5 were broken.Fragments of the broken part remained in the patient.Back pain was reported as the result of the event.A primary surgery was performed due to the event.
 
Additional Manufacturer Narrative
Product analysis: visual examination of the mas bone screw confirmed bone screw complete fracture at approximately ~4 threads from the base of the bone screw head.Microscopic examination did not identify material defect near the area of fracture origin, which could contribute to crack propagation.Microscopic examination of the fracture surface identified gently curving convex striations through the cross-sectional area of the implant, which is indicative of cyclic fatigue until subsequent mechanical failure of the screw.Dimensional inspection of major diameter confirms conformance to print specification.After visual, optical and dimensional inspection, no evidence was found that would suggest a defect in manufacturing of the implant components.The above observations are consistent with cyclic fatigue.Due to imdrf harmonization, some previously submitted device, method, result, and conclusion codes related to this event may have been updated.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key7404857
Report Number1030489-2018-00524
Device Sequence Number14006013
Product Code KWP
UDI-Device Identifier00613994972019
UDI-Public00613994972019
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K113174
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number55840005550
Device Lot NumberH5315433
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/25/2018
Initial Date Received by Manufacturer 03/09/2018
Supplement Date Received by Manufacturer06/26/2018
08/08/2018
Initial Report FDA Received Date04/06/2018
Supplement Report FDA Received Date07/26/2018
10/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-