|
It was reported five ultrathane mac-loc locking loop multipurpose drainage catheters fell out on five different patients¿ in prior two weeks.As reported, all five devices were from the same lot and three of the five were inserted by senior consultants.To date, additional information regarding event dates, patient demographics and patients¿ has been requested but the status is unknown.The following medwatch reports are associated: 1820334-2018-00863, 1820334-2018-01010, 1820334-2018-01011, 1820334-2018-01012, and 1820334-2018-01013.
|
|
Additional information received (11jun2018) identified that per the representative, no damage was noted to the drainage catheters upon observation.No further information is available.Investigation - evaluation: a review of the complaint history, device history record, documentation, drawing, instructions for use (ifu), manufacturing instructions, quality control, and specifications of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.Document review has established that proper procedures are in place to identify/prevent this failure mode prior to distribution.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were five other reported complaints for this lot number.All complaints were reported for the same failure and from the same customer.The instructions for use (ifu) was reviewed and noted two steps that are designed to help keep the catheter secure within the patient.Information was requested but not provided on whether the proper procedures were followed correctly.The ifu outlines procedures to follow for correct use of the catheter in addition to showing diagrams on how to properly pull the suture and lock the mac-loc lever.The document states the following instructions: "stabilize the mac-loc catheter hub assembly with one hand and pull back on the drawstring to form the distal catheter loop configuration.(fig.1)." "while maintaining traction on the drawstring, push the locking cam lever down until a distinct "snap" is felt.The distal loop of the catheter is now locked into position.(fig.2)." based on the information provided, no product returned and the results of our investigation, the likely root cause has been determined to be product use or handling related (user technique).We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment no further action is required.
|