| Model Number |
MMT-751PNAB |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had a blank display.The customer's blood glucose level was 142 mg/dl.The customer was assisted with the troubleshooting.The insulin pump will be returned for the analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Pump passed the operating currents, self test and displacement test.No blank display anomaly noted during test.
|
| |
|