Model Number MSERIES CCT
Medical Device Problem Code Loss of Power (1475)
Health Effect - Clinical Code No Information (3190)
Date of Event 03/23/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The malfunction was observed during initial testing; however after disassembling the device to determine root cause, the reported malfunction was no longer seen.The device was put through extensive testing without duplicating the reported malfunction.The battery interconnect board was replaced as a precaution.The device successfully passed the final test procedure, was recertified, and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
 
Event or Problem Description
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device inappropriately shut down.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M SERIES
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key7406121
Report Number1220908-2018-00996
Device Sequence Number19145595
Product Code MKJ
UDI-Device Identifier00847946004828
UDI-Public00847946004828
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K011865
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMSERIES CCT
Device Catalogue NumberM SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/26/2018
Initial Report FDA Received Date04/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-