• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO 3.0MM DMND MTCH HD, 16CM; DRILLS, BURRS, TREPHINES & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO 3.0MM DMND MTCH HD, 16CM; DRILLS, BURRS, TREPHINES & ACCESSORIES Back to Search Results
Model Number 8470107030D
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is available for return.A follow up report will be filed once the quality investigation is complete.Awaiting device return.
 
Event or Problem Description
It was reported that when performing spinal surgery, the bur broke into two pieces.It was also reported that there were no delays and no adverse consequences as a result of this event.It was further reported that the procedure was completed successfully.
 
Additional Manufacturer Narrative
Investigation results indicate that the bur experienced a high axial force which increased the stress at the mid bearing causing the bur to fracture at the location where heat and wear developed.The ifu for attachments associated with these cutting accessories warns the user against this type of use: "do not apply excessive pressure, such as bending or prying, with the cutting accessory.Excessive pressure may bend or fracture the cutting accessory." the quality investigation is complete.
 
Event or Problem Description
It was reported that when performing spinal surgery, the bur broke into two pieces.It was also reported that there were no delays and no adverse consequences as a result of this event.It was further reported that the procedure was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.0MM DMND MTCH HD, 16CM
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key7406471
Report Number0001811755-2018-00574
Device Sequence Number173448
Product Code HBE
UDI-Device Identifier07613327097214
UDI-Public(01)07613327097214
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K143540
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/01/2021
Device Model Number8470107030D
Device Catalogue Number8470107030D
Device Lot Number16019017
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/10/2018
Initial Date Received by Manufacturer 03/14/2018
Supplement Date Received by Manufacturer03/14/2018
Initial Report FDA Received Date04/09/2018
Supplement Report FDA Received Date05/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No