| Model Number |
295251 |
| Medical Device Problem Code |
Cut In Material (2454)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Upon triage the service tech found the pump to have a damaged fan and a damaged power cord with exposed copper.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation of the scd700 was performed and the customer¿s reported condition was confirmed and it was noted the unit had a damaged power cord with exposed copper.The device was cleaned and disinfected according to manufacturer's guideline.The power cord has exposed copper wires and was replaced.The device was fully checked for functionality and patient electrical safety according to manufacturer's guideline.The potential root cause is customer misuse by running over the cord with hospital bed or due to the procedure used to unplug the unit and the procedure of wrapping of the cord around the bed hook.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
|
| |
|