Model Number CSK-TC10-3M
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/22/2018
Type of Reportable Event Malfunction
Event or Problem Description
Device analysis was performed on the returned electrode and found that the electrode shaft was separated from the hub and not returned for analysis.Root cause of the failure is excessive external mechanical force placed upon the shaft which led to the complete separation of the shaft from the hub.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CSK-TC ELECTRODE
Common Device Name
PROBE, RADIOFREQUENCY LESION
Manufacturer (Section D)
COSMAN MEDICAL
22 terry ave
burlington MA 01803
MDR Report Key7406973
Report Number3006630150-2018-01283
Device Sequence Number1007482
Product Code GXI
UDI-Device Identifier00813250011524
UDI-Public00813250011524
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K060799
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 04/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model NumberCSK-TC10-3M
Device Lot NumberL486
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/06/2018
Initial Report FDA Received Date04/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-