Model Number E051
Device Problems Premature Discharge of Battery (1057); Device Displays Incorrect Message (2591)
Patient Problem Pneumonia (2011)
Event Date 11/22/2017
Event Type  Injury  
Manufacturer Narrative
At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.
 
Event Description
Boston scientific received information that this patient presented to the emergency room with pneumonia.Upon interrogation of the device, it was found that there was a recorded code 1003 indicative of battery voltage too low for the projected remaining capacity.Data analysis showed the battery appeared to be depleting more quickly than expected.Device replacement was recommended.This device was explanted.No additional adverse patient effects were reported.
 
Manufacturer Narrative
This device was thoroughly inspected and analyzed upon receipt at our quality assurance laboratory.Review of the device memory indicated that a low voltage alert, code 1003, was recorded.The battery voltage was lower than expected.Using historical daily battery voltage measurement data, engineers determined that this device was demonstrating behavior consistent with a high current condition associated with a compromised low voltage capacitor connected to the device's battery.Low voltage capacitors are used in the device's high voltage charging operation in order to facilitate fast charge times.Malfunction of these capacitors resulted in a high current drain, which was depleting this device's battery faster than normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUNCTUA
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7407198
MDR Text Key104699293
Report Number2124215-2018-01574
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 03/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/09/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/21/2012
Device Model NumberE051
Other Device ID NumberPUNCTUA ICD
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received03/14/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/23/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
0095; 1742; 1790; 1861; E051
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age71 YR
-
-