| Model Number |
1516-20-708 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Muscular Rigidity (1968); Pain (1994); Loss of Range of Motion (2032); Limited Mobility Of The Implanted Joint (2671); Not Applicable (3189); No Code Available (3191)
|
| Date of Event |
03/12/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Patient was revised for painful total knee and limited range of motion.Components were found to be well fixed.Parts were removed in favor of a revision knee.Patient consequence? :no.Is the information being submitted for this complaint all the details that are known/available regarding this event? : yes.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).(b)(4) is used to capture the surgical intervention and medical device removal.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Medical record ad 17 june 2019 was reviewed on 9 december 2019.(b)(6) 2017: the patient underwent a right total knee arthroplasty secondary to osteoarthritis and pain.Attune implants were used with depuy cement x1.The patella was resurfaced.No intraoperative complications were noted.(b)(6) 2018: the patient underwent a revision of the right knee due to pain stiffness.Intraoperatively, femoral component was found to be internally rotated ¿slightly¿ and noted as a contributor to the issue.All components were replaced with non-depuy revision system with non-depuy cement.No intraoperative complications during revision surgery.Doi: (b)(6) 2017; dor: (b)(6)2018: (patellar, tibial, insert, femoral).
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|