|
WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.7MM TI CORTEX SCREW SLF-TPNG WITH T8 STARDRIVE RECESS 20MM; SCREW, FIXATION, BONE
|
Back to Search Results |
|
| Model Number |
402.880 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient height reported as 5¿11¿.Additional product code: hrs.Date of implant reported as (b)(6) 2014, exact date is not known.Date returned to manufacturer.Subject device has been received and is currently in the evaluation process.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported patient was implanted with a locking compression plate (lcp) wrist fustion short bend plate and eight (8) screws in (b)(6) 2014.The area had fused and healed completely.It was subsequently determined that the plate and two (2) screws had broken.Patient was returned to surgery on (b)(6) 2018 for removal of all hardware.All hardware, with the exception of the tips of the two (2) broken screws, were removed successfully.The broken tips remain embedded in the patient¿s bone.No patient consequence was reported.Product was identified upon receipt at manufacturer location, device added to this report.Concomitant devices reported: 2.7mm cortex screw 14mm (402.874, lot 9096654, quantity 1), 3.5mm locking screw 20mm (412.106, lot 9054496, quantity 1), 3.5mm locking screw 28mm (412.110, lot 8634026, quantity 1), 3.5mm stardrive cortex screw 20mm (04.200.020, lot 8844355, quantity 1), 3.5mm stardrive cortex screw 18mm (04.200.018, lot number unknown, quantity 1), screw (part number unknown, lot number unknown, quantity 1).This report is for one (1) 2.7mm cortex screw 20mm this is report 3 of 3 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Dhr review was completed.Part: 402.880, lot: 9141111, manufacturing site: bettlach, release to warehouse date: 05.Sep.2014, no ncrs were generated during production.Review of the device history record showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Customer quality conducted an investigation of the returned device.Device condition: the plate (part 402.880, lot 9141111, mfg 05.Sep.2014) was received at us cq broken.The complaint condition was visually confirmed and consistent with the reported condition.Replication of the complaint condition is not applicable as the devices were returned damaged.Lot number review: dhr review showed no ncr¿s were generated during production.The material, and material properties of the returned part were determined to be conforming at the time of manufacture based on review of the dhr(s).Drawing review: the drawing was reviewed; dimensional analysis was not performed as relevant features are significantly deformed.During the investigation no product design issues or discrepancies were observed that may have contributed to the complaint condition.Conclusion: the complaint condition is confirmed as the plate was received broken.A device inspection, device history record review and drawing review were performed as part of this investigation.There is no indication that a design or manufacturing issue contributed to the complaint.While no definitive root cause could be determined it is possible that the device encountered unintended forces prior to bone healing.No new malfunctions were observed during the course of this investigation.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a hardware removal procedure on (b)(6) 2018 due to broken hardware.A titanium locking compression (lcp) wrist fusion plate and eight screws were originally implanted in the patient on (b)(6) 2014 as part of a right wrist total arthrodesis.Following the original procedure the patient attended four follow up appointments (b)(6) 2015.At the first appointment on (b)(6) 2014 the patient was placed in a short arm cast, and a right wrist ap/lat was taken, as well as x-rays (not available).The ap/lat revealed that the fusion was intact and the hardware were stable.At the second appointment on (b)(6) 2015 the cast was removed and additional x-rays and an right wrist ap/lat revealed that bony healing was present and stable hardware.At that time the patient was still experiencing some numbness in the dorsal hand around the incisions as well as, some tightness in the knuckles.The patient denied feeling any pain at that time.A right wrist ap/la taken at the third appointment on (b)(6) 2015 revealed the development of a right fusion mass.The hardware was still stable, with the patient experiencing little to no pain.Upon examination the patient¿s incision was well healed as well as, full motion of finger range.There was mild dorsal and hand swelling.The patient was told to refrain from heavy lifting.At the fourth appointment ((b)(6) 2015) x-rays revealed the hardware was still intact and fusion mass was still developing.The patient at that time was reported to be doing very well.The patient was no longer experiencing any pain and had begun to engage in normal activities.At that point in time it had been 2 weeks since the patient last wore his brace.The patient was still told to reframe from heavy lifting for another month.However, despite the patients promising post-operative progression, it was subsequently discovered that the patient had only experienced fusion at the radiocarpal and midcarpal joints and not at the third cmc joint and some of the hardware had broken.All of the hardware, except for the shafts of two screws that were encased in the bone of the third metacarpal were successfully removed during the (b)(6) 2018 procedure.There was no reported patient consequence.Concomitant devices reported: 2.7mm cortex screw 14mm (402.874, lot 9096654, quantity 1), 3.5mm locking screw 20mm (412.106, lot 9054496, quantity 1), 3.5mm locking screw 28mm (412.110, lot 8634026, quantity 1), 3.5mm stardrive cortex screw 20mm (04.200.020, lot 8844355, quantity 1), 3.5mm stardrive cortex screw 18mm.
|
| |
|
Search Alerts/Recalls
|
|
|