Catalog Number 114816
Device Problem Fracture (1260)
Patient Problem No Code Available (3191)
Event Date 03/06/2018
Event Type  Injury  
Event Description
Revision surgery - received an email from zimmer biomet on (b)(6) 2018, stating that, "it was reported that the patient underwent an initial left total elbow procedure on (b)(6) 2010.Subsequently, the patient was revised due to implant fracture on (b)(6) 2018.
 
Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was due to an implant fracture.The previous surgery and the revision detailed in this investigation occurred 7.6 years apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the reported event.As of 4-may-2018, the records needed, to verify an acceptable sterilization process, have not been forwarded by zimmer-biomet.Should zimmer-biomet provide the needed records at a later time, this investigation shall be updated.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to implant fracture.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Manufacturer Narrative
Manufacturer narrative: the reason for this revision surgery was due to an implant fracture.The previous surgery and the revision detailed in this investigation occurred 7.6 years apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the reported event.As of 4-may-2018, the records needed, to verify an acceptable sterilization process, have not been forwarded by zimmer-biomet.Should zimmer-biomet provide the needed records at a later time, this investigation shall be updated.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to implant fracture.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISCOVERY ELBOW
Type of Device
DISC ULNA 3X115MM LT W BRNG C
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer Contact
teffany hutto
9800 metric blvd
austin, TX 78758-5445
MDR Report Key7408212
MDR Text Key104730423
Report Number1644408-2018-00305
Device Sequence Number1
Product Code JDC
UDI-Device Identifier00888912225168
UDI-Public(01)00888912225168
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K013042
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 11/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/09/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2015
Device Catalogue Number114816
Device Lot Number332520
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/28/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
114700 LOT 082120.; 114906 LOT 836600.; 402283 LOT 504960.; 402283 LOT 504960.; 414991 LOT 715060.
Patient Outcome(s) Required Intervention;
Patient Age60 YR
Patient SexMale
-
-