Catalog Number PTA5-35-80-6-8.0
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that during an unspecified procedure, the advance 35 lp low profile balloon catheter's balloon ruptured at 15 bar when inflated.Subsequently, the advance 35 lp low profile balloon catheter was removed and another manufacturer's device was used to complete the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Additional questions regarding details of the event have been asked, but are not available at this time.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: a review of the complaint history, device history record, documentation, drawing, instructions for use (ifu), quality control, specifications, and a visual inspection of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed that the device was returned in a damaged state.The balloon was inflated to 10 atmospheres (atm), and leakage was visible at the distal end of the catheter.A kink in the balloon tubing was noted at a distance of 5.7 cm from the distal end of the device.The balloon tubing was also twisted, likely due to the observed kink.The catheter tubing was cut open proximal to the proximal bond, and a rupture was noted inside the lumen wall.There were no balloon defects or damage noted; the balloon did not burst.Additionally, a document based investigation evaluation was performed.A review of the device history record confirms that the lot was released meeting all finished goods release criteria.There is no evidence to suggest that there is any nonconforming product in the field.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the quality engineering risk assessment, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 35 LP LOW PROFILE BALLOON CATHETER
Common Device Name
DQY CATHETER, PERCUTANEOUS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7408247
Report Number1820334-2018-00833
Device Sequence Number1252121
Product Code DQY
UDI-Device Identifier10827002523477
UDI-Public(01)10827002523477(17)200413(10)7828532
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K091527
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPTA5-35-80-6-8.0
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 03/19/2018
Supplement Date Received by Manufacturer06/01/2018
Initial Report FDA Received Date04/09/2018
Supplement Report FDA Received Date06/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-