The reported event that bone screw, t7, diam.2.7x14mm was alleged of issue s-11 (breakage during surgery) could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.The batch record could not be reviewed because the affected device was not returned and the lot number was not communicated.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.However, based on complaint history, the potential scope of application of excessive force (torsional) during screw insertion leading to the breakage of the screws cannot be excluded.Thus, the root cause can most likely be attributed to a user related issue rather than any deficiency in the product.A review of the labeling did not indicate any abnormalities.The instruction for use clearly mentions that ¿screws should not be over-tightened during insertion.Excessive over-tightening will compromise the integrity of the screw head, result in possible screw and screwdriver blade breakage or deformation, and lead to loss of friction fit performance¿.The instruction for use also suggests that if the insertion of the 2.7 mm screw is hindered, it is recommended to use the 2.7 mm tap (62-27010) or the 2.3 mm cortical drill bit (60-23141, 60-23341, 60-23441) to reduce the risk of screw breakage during insertion.Thus, failure to follow the operative technique and the instructions by the user is the root cause for the screw head breakage.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|