• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MEDICAL PRODUCTS KENDALL SCD; SLEEVE, LIMB, COMPRESSIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MEDICAL PRODUCTS KENDALL SCD; SLEEVE, LIMB, COMPRESSIBLE Back to Search Results
Model Number 29525
Medical Device Problem Code Cut In Material (2454)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The customer reports the pump has a damaged bed hook and the power cord is damaged with exposed copper.
 
Additional Manufacturer Narrative
An evaluation of the scd 700 was performed and the customer states ¿ the pump has a damaged bed hook and the power cord is damaged with exposed copper¿ the unit was triaged and the customer¿s reported condition was confirmed.The power cord was replaced.The potential root causes are customer misuse due to the procedure used to unplug the unit and the procedure of wrapping of the cord around the bed hook.A device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KENDALL SCD
Common Device Name
SLEEVE, LIMB, COMPRESSIBLE
Manufacturer (Section D)
COVIDIEN MEDICAL PRODUCTS
building 10- no 789 puxing roa
shanghai 20111 4
CN  201114
MDR Report Key7409829
Report Number3006451981-2018-00268
Device Sequence Number2023131
Product Code JOW
UDI-Device Identifier10884521129566
UDI-Public10884521129566
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/23/2023
Device Model Number29525
Device Catalogue Number29525
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2018
Initial Date Received by Manufacturer 04/05/2018
Supplement Date Received by Manufacturer04/05/2018
Initial Report FDA Received Date04/09/2018
Supplement Report FDA Received Date04/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-