MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problems No Display/Image (1183); Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 03/20/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A manufacture representative inspected the navigation system onsite and confirmed the reported issue.When turning on the navigation system from an off position, the system would appear on, however the staff cart screen showed no signal.The representative removed the old computer, installed a new one that has digital visual interface (dvi), and video graphics array (vga) ports.All connections were checked, and when powering up there was no signal.The representative attempted to plug the video adapter into itself without resolution.The new computer was removed and the original was put back in.The cords were arranged and the computer operated effectively.All applications were tested and performed as expected.The computer was returned to the manufacturer for evaluation.No fault was found with the returned computer.All tests passed and no errors were found.
 
Event Description
Information was received from a manufacturer representative regarding a navigation system outside of a procedure.It was reported that the monitors were flickering.There was no patient present when this issue was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
stephanie riley
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7409850
MDR Text Key104841142
Report Number1723170-2018-01555
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 04/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Model NumberS7
Device Catalogue Number9733856
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received03/20/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/19/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-