Model Number U3575615
Medical Device Problem Codes Retraction Problem (1536); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the sfa ,using an antegrade approach, access was gained in the common femoral artery.Reportedly, the pta balloon allegedly was not able to be retracted through the 5fr sheath and subsequently detached inside the sheath.The balloon and sheath were removed together as one unit.Another balloon and sheath were used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.Investigation summary: the device was not returned, however five electronic photos were provided for review.The photos show a segment of the device inserted into the user's introducer sheath with bunching noted to the distal balloon material.The distal tip of the sheath was also noted to be flared.A complete detachment was also noted in the outer catheter shaft proximal to the proximal balloon weld joint.Therefore, based on the photo review, the investigation can be confirmed for the reported retraction issue and catheter detachment.It is possible that the identified retraction issues led to the catheter detachment, if excessive force was applied to the device during removal.However, the definitive root cause for the identified balloon detachment or retraction issue could not be determined based upon the available information.It is unknown if patient and/or procedural issues contributed to the reported event.Labeling review: the current ifu (instructions for use) states: warnings: do not exceed the rbp recommended for this device.Balloon rupture may occur if the rbp rating is exceeded.To prevent over pressurization, use of a pressure monitoring device is recommended.Precautions: if resistance is felt during post procedure withdrawal of the catheter through the introducer sheath/guide catheter, determine if contrast is trapped in the balloon with fluoroscopy.If contrast is present, push the balloon out of the introducer sheath/guide catheter and then completely evacuate the contrast before proceeding to withdraw the balloon.If resistance is still felt during post procedure withdrawal of the catheter, it is recommended to remove the balloon catheter and introducer sheath/guide catheter as a single unit.Potential adverse reactions: additional intervention.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the sfa, using an antegrade approach, access was gained in the common femoral artery.Reportedly, the pta balloon allegedly was not able to be retracted through the 5fr sheath and subsequently detached inside the sheath.The balloon and sheath were removed together as one unit.Another balloon and sheath were used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 035 PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7409954
Report Number2020394-2018-00359
Device Sequence Number19136573
Product Code LIT
UDI-Device Identifier00801741093005
UDI-Public(01)00801741093005
Combination Product (Y/N)N
PMA/510(K) Number
K142261
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2020
Device Model NumberU3575615
Device Catalogue NumberU3575615
Device Lot NumberCMBW0372
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/15/2018
Supplement Date Received by Manufacturer06/06/2018
Initial Report FDA Received Date04/09/2018
Supplement Report FDA Received Date07/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age75 YR
Patient Weight70
-
-