• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Catalog Number RONYX22530UX
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Patient Problem/Medical Problem (2688)
Date of Event 03/16/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An attempt was made to use one resolute onyx drug eluting stent to treat a severely tortuous and severely calcified lesion, located in the mid diagonal branch, exhibiting 100% stenosis (cto).There was no damage noted to packaging.There were no issues noted when removing the device from the hoop.There were no difficulties removing the protective sheath.The device was inspected with no issues identified.The lesion was pre-dilated.There was no issue with the inflation device.The device did pass through a previously-deployed stent.Resistance was not encountered when advancing the device.Excessive force was not used during delivery.It was reported that the stent dislodgement occurred during delivery.It was reported that the device went over a bend and when it was backed up, the stent dislodged, and a non-compliant post dilation balloon was then used to fully expand the stent to the vessel wall.The stent was not crushed.The procedure was completed.Patient status post-procedure is alive with no injury.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7410610
Report Number9612164-2018-00671
Device Sequence Number1373402
Product Code NIQ
UDI-Device Identifier00643169556751
UDI-Public00643169556751
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/29/2019
Device Catalogue NumberRONYX22530UX
Device Lot Number0008772717
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/16/2018
Supplement Date Received by Manufacturer08/03/2018
Initial Report FDA Received Date04/10/2018
Supplement Report FDA Received Date08/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-