| Model Number |
ENVEOR-N-US |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/19/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned, therefore no product analysis can be performed.Without return of the device, a conclusive cause cannot be determined.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that during the implant of this transcatheter bioprosthetic valve with this delivery catheter system (dcs), the valve was deployed and recaptured four times successfully.During the fifth deployment attempt, the capsule was not retracting, and a capsule separation occurred.The system was removed carefully through arteriotomy causing no injury to the vessel.A new valve and dcs was prepped and successfully deployed.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received reported that the valve was deployed and recaptured three times successfully.During the fourth deployment attempt, the capsule was not retracting, and a capsule separation occurred.The valve was fully recaptured prior to removal.It was noted that a steep root angle was encountered, which required a fair amount of torque on the catheter.No additional adverse patient effects were reported.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at medtronic's quality laboratory, the valve was received loaded in the capsule of the delivery catheter system (dcs).The device was received with the capsule partially open.The handle appeared intact.The trigger moved to fully advance and retracted positions and locked in place when released.The tip-retrieval mechanism appeared intact.The device was returned with the end cap/screw gear snap fit connected.Several voids were observed over the nitinol reinforcing frame along the length of the capsule.There was a break observed in the capsule nitinol reinforcing frame near the proximal end of the capsule.The separation site was jagged and uneven.The capsule appeared bent or bowed.The valve cannot be removed from the dcs.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: the device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.It was reported that the valve was placed at a suboptimal position, was recaptured three times, and deployment was attempted four times.Various factors can affect valve positioning, such as patient anatomy or physician experience; the cause of the suboptimal placement could not be determined with the limited information available.The recapture is a feature of the enveor device that allows for additional attempts at accurately positioning the valve; however, the device instructions for use (ifu) instruct the user to a maximum of three recaptures, which then warrants withdrawal of the valve from the patient.In this case, the valve was deployed a fourth time against ifu recommendations, and a capsule separation occurred warranting withdrawal of the device.It is possible that the fourth deployment attempt may have resulted in excessive force on the system leading to the capsule separation, in addition to the reported aortic root angulation.However, without review of procedural images, the series of events leading to the capsule separation cannot be assessed, and an assignable root cause cannot be determined.Evaluation of the returned device confirmed the capsule separation; several voids were observed over the nitinol reinforcing frame along the length of the capsule, and a break was observed in the capsule nitinol reinforcing frame near the proximal end of the capsule.The separation site was jagged and uneven.The capsule appeared bent or bowed.The valve could not be removed from the delivery catheter system (dcs).These observations on the returned dcs indicate delamination between the capsule outer polymer and the nitinol frame, and typically occur when the capsule is subjected to a bending force potentially after tracking through tortuous anatomy, or if a valve was misloaded.Capsule separation typically occurs due to excessive compressive forces applied during the valve loading process.If information is provided in the future, a supplemental report will be issued.
|
| |
|