| Model Number |
4543 |
| Medical Device Problem Codes |
Fracture (1260); High impedance (1291); Capturing Problem (2891)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/06/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The product has been received for analysis.This report will be updated upon completion of analysis.
|
| |
|
Event or Problem Description
|
|
Boston scientific received information that this left ventricular (lv) lead had observations of high pacing thresholds and out of range high pacing impedances greater than 2000 ohms.The lead was explanted and replaced, where it was observed that the lead was fractured.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, this lead was confirmed to be fractured 35.8 cm from the terminal pin.Based on the analysis results, we suspect that fatigue or stress in the region of the fracture site due to relative motion between the suture sleeve and an anatomical feature led to the fracture.As a lead moves in response to normal heart rhythms and blood flow, extensive flexing over a period of time may cause fatigue or stress, weakening the coil, ultimately resulting in a fracture.This can occur between the suture sleeve and some other part of the anatomy.Fatigue fractures in the pocket or clavicular/first rib area are well known and documented in the industry.A combination of lead design, implant techniques, and patient anatomy and activity level contribute to these types of occurrences.Laboratory analysis confirms that the allegations of high impedance and thresholds were likely due to the conductor damage in the area of the suture sleeve tie-down.
|
| |
|
Search Alerts/Recalls
|