| Model Number |
D314DRG |
| Medical Device Problem Codes |
Battery Problem (2885); Data Problem (3196)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/25/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product event summary: performance data collected from the device was received and analyzed.Analysis of the device memory indicated pacing capture threshold in the left ventricle was high.The device was returned and analyzed.Analysis of the device revealed normal battery depletion.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead had a high threshold.The physician felt that capture management of the implantable cardioverter defibrillator (icd) may have recorded the threshold high than normal.The lead was reprogrammed.It was additionally reported the device triggered an alert for recommended replacement time (rrt) a few days later.The device was explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|