As reported to angiodynamics on march 28, 2018: at the start of an endovenous laser treatment procedure, it was noted the laser fiber was bent/fractured near the tip.The device was set aside and a new of the same was used to successfully complete the procedure.There was no patient harm or injury as the device had not come into contact with the patient.It was indicated the disposable device is available for return to the manufacturer for a device evaluation.
|
|
Returned for evaluation was an evlt/400micron fiber.A visual inspection noted that 88.2cm from the tip, the fiber is fractured.The customer's reported complaint description of a broken (cracked) fiber is confirmed.A root cause for the complaint description cannot be definitively determined.A review of the lot history records was performed for the reported packaging lot for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.At the vendor facility, during the manufacturing process, fibers are repeatedly bent and coiled and there is a final visual inspection performed with a hene laser to check the entire fiber length for defects prior to shipment.During the packaging step, the fibers are inspected for damage.It is unlikely that the fiber was fractured prior to packaging.The most likely root cause to this event is due to handling after the device left the angiodynamics facility.The instructions for use, which is supplied to the end user with this catalog number contains the following statement "prior to and during use, avoid damaging the fiber by striking, stressing or excessive bending.Do not coil the fiber tighter than a radius of 60mm".A review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference complaint #(b)(4).
|