• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS ANGIOOPTIC ANGIOGRAPHIC CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS ANGIOOPTIC ANGIOGRAPHIC CATHETER Back to Search Results
Catalog Number H787106322015
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported defective device has been returned to the manufacturer and an investigation into the root cause for event is currently in progress.The results of the device evaluation will be sent via a follow up medwatch.(b)(4).
 
Event or Problem Description
As reported to angiodynamics on march 13, 2018: during an angiographic procedure, when threading the catheter over the guidewire, it was noted the tip of the catheter had cracked.There was no report of any patient harm or injury due to this event.The reported defective disposable device has been returned to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
Returned for evaluation was one angiographic catheter, p/n: 10632201.Lot number 5275911.A visual examination of the returned device noted two kinks in the catheter shaft.It was also noted the tip of the catheter was damaged/fractured 4.2cm from the tip end.The returned device met all manufacturing dimensional specifications for this product.The customer's reported complaint description of the catheter tip being cracked/fractured is confirmed.Although the complaint description is confirmed, a definitive root cause cannot be determined.However, it does not appear to be manufacturing related.The instructions for use, which is supplied to the end user with states; "tip straighteners are furnished on all catheters intended to be straightened with the aid of a tip straightener.Reshaping of the catheter tip is not recommended.Physical damage to the catheter material can result when exposed to heat.Never advance or retract an angiographic catheter or guidewire against resistance.This may result in damage to the vessel, the product, or both." the label indicates a recommended guidewire size of.035".A review of the lot history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.Manufacturing personnel are directed to 100% visually inspect this product to verify there are no pin holes or cuts to the packaging, and there is no component damage.This type of defect would be noticed prior to shipment.A review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOOPTIC ANGIOGRAPHIC CATHETER
Common Device Name
ANGIOOPTIC ANGIOGRAPHIC CATHETER
Manufacturer (Section D)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
MDR Report Key7411794
Report Number1319211-2018-00033
Device Sequence Number19148372
Product Code DQO
UDI-Device IdentifierH787106322015
UDI-PublicH787106322015
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K112452
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 07/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date11/30/2020
Device Catalogue NumberH787106322015
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2018
Supplement Date Received by Manufacturer03/13/2018
Initial Report FDA Received Date04/10/2018
Supplement Report FDA Received Date07/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-