U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP CRYOCONSOLE; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP CRYOCONSOLE; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number 106A3
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/20/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during a cryo ablation procedure, a system notice was received indicating a mechanical component error.Additionally, a system notice was received indicating that there is a problem with the system.It was also reported that flow and pressure issues were observed.The console was rebooted several times without resolve.The case was aborted, and the patient was under general anesthesia.A field service visit was recommended.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the data files were returned and analyzed.The data files confirmed system notice 50022 (mechanical component error) was triggered four times on the date of the event.A pressure transducer (console part) was returned and serviced.Visual inspection showed the product was intact with no apparent issues.The pressure transducer passed the performance test as per specification.The product issues reported (system notice 50022 and 11200) are not likely to cause or contribute to a death or serious injury; however, the risk of the patient being under general anesthesia without full therapeutic effect is the adverse event being reported.The decision to abort the procedure without use of alternate therapy was based upon the medical judgment of the physician.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRYOCONSOLE
Common Device Name
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
MDR Report Key7412331
Report Number3002648230-2018-00209
Device Sequence Number16452900
Product Code LPB
UDI-Device Identifier00613994686077
UDI-Public00613994686077
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P020045
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number106A3
Device Catalogue Number106A3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2018
Initial Date Received by Manufacturer 03/20/2018
Supplement Date Received by Manufacturer04/18/2018
Initial Report FDA Received Date04/10/2018
Supplement Report FDA Received Date05/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-