|
|
| Catalog Number |
4002C0830 |
| Medical Device Problem Codes |
Failure to Advance (2524); Physical Resistance (2578)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Results: the pet lock was intact on the proximal end of the pusher assembly.The embolization coil was intact with its pusher assembly and had offset coil winds.The pusher assembly mid-joint was retracted approximately 26.5 from proximal to the introducer sheath.Conclusions: evaluation of the returned device revealed that the embolization coil had offset coil winds.If the device is forcefully advanced against resistance, damage such as this may occur.The root cause of the resistance could not be determined.Further evaluation of the returned device revealed that the pusher assembly mid-joint was retracted proximal to the introducer sheath friction lock.This damage likely occurred after the procedure.Evaluation of the returned device revealed that the device was kinked.If the device is forcefully advanced against resistance, damage such as this may occur.The root cause of resistance could not be determined.Further evaluation of the returned device revealed that the sr wire was fractured and the embolization coil was detached from its pusher assembly.If the proximal end of the embolization coil is forcefully retracted against resistance through the introducer sheath friction lock, damage such as this may occur.The introducer sheath¿s friction lock id is smaller than the rest of the sheath.The id of the friction lock is slightly larger than the proximal end of the embolization coil.However, if the coil is retracted through the friction lock, resistance may be encountered, and will likely result in a sr wire fracture.Evaluation of the returned handle revealed that the handle was functional.The handle was functionally tested on the handle fixture for pull force and was within specification.The lantern referred to in the complaint was returned to penumbra for evaluation.Penumbra coils and catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2018-00745.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the left internal iliac artery using penumbra coils 400 (pc400¿s).It was noted that the patient¿s anatomy was tortuous.During the procedure, while attempting to advance a pc400 through a lantern delivery microcatheter (lantern), the physician experienced resistance approximately 10 centimeters from the distal tip of the lantern and the pc400 would not advance; therefore, it was removed.The physician then attempted to advance a new pc400 through lantern; however, resistance was encountered at the same position and pc400 would not advance further.Therefore, the pc400 was removed and the procedure was completed using other coils and the same lantern.There was no report of an adverse effect to the patient.
|
| |
|
Search Alerts/Recalls
|
|
|