• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. VENTRALIGHT ST W/ ECHO; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. VENTRALIGHT ST W/ ECHO; SURGICAL MESH Back to Search Results
Catalog Number 5955800
Medical Device Problem Codes Shelf Life Exceeded (1567); Use of Device Problem (1670)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As reported the expired device was inadvertently implanted.Our review shows that the lot was provided to the customer prior to the expiration date.The product is packaged with the expiration clearly displayed on each level of packaging.The instructions-for-use instructs the user to inspect the package prior to use.This event is confirmed to be solely a use related error.Note: the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.Implanted.
 
Event or Problem Description
It was reported that on (b)(6) 2018 during a laparoscopic repair of an incarcerated ventral hernia a davol ventralight st w/ echo was implanted that had been retrieved from the hospital inventory.As reported, after the procedure was completed, during a routine documentation of the devices used in the procedure it was noted the expiration date of the mesh was (b)(6) 2018.As reported the surgeon decided to leave the mesh implanted with no further medical or surgical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTRALIGHT ST W/ ECHO
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC.
100 crossings blvd.
warwick RI 02886
MDR Report Key7412924
Report Number1213643-2018-00986
Device Sequence Number19194095
Product Code FTL
UDI-Device Identifier00801741031779
UDI-Public(01)00801741031779
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K130968
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/28/2018
Device Catalogue Number5955800
Device Lot NumberHUAN0488
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2018
Initial Report FDA Received Date04/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-