ETHICON INC. PDS II SUTURE 27"(70CM) 2-0 VIO; SUTURE, SURGICAL, ABSORBABLE Back to Search Results
Catalog Number Z333H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported suture was used during a veterinarian procedure.The needle broke approximately 2 mm above the swage.
 
Additional Manufacturer Narrative
The product upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
 
Additional Manufacturer Narrative
(b)(4).Date sent to the fda: 05/08/2018 device evaluation summary: representative samples were returned for evaluation.The sample was opened and the swage and attachment area were as expected; also, the tip and body of the needle no defects or needle breakage was observed.They were visually and functionally examined for resistance needle strength and they met the requirements.Per the condition of the representative samples, no performance ¿ breakage needle were found on the samples and the tested sample met the requirements.The actual device was returned for evaluation.A fracture was observed at the suture attachment of the needle.The needle was received in two pieces, only one of the mating fracture surfaces was microscopically examined for this evaluation.The evaluation revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile overload failure.A cup-and-cone shaped fracture and macroscopic plastic deformation observed coincidental to the fracture provide additional evidence that the failure was induced by mechanical deformation leading to ductile overload.This was a ductile fracture.The evidence of this examination indicates that the breakage occurred at the attachment area of the needle during use due to tensile overload.There is no evidence of any material flaw or defect that would cause premature failure.In order to avoid this kind of damage grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the attachment end to the point.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PDS II SUTURE 27"(70CM) 2-0 VIO
Common Device Name
SUTURE, SURGICAL, ABSORBABLE
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
MDR Report Key7413157
Report Number2210968-2018-72047
Device Sequence Number16418219
Product Code NEW
UDI-Device Identifier10705031060548
UDI-Public10705031060548
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
N18331
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Catalogue NumberZ333H
Device Lot NumberLG6847
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2018
Initial Date Received by Manufacturer 03/19/2018
Supplement Date Received by Manufacturer04/04/2018
05/02/2018
Initial Report FDA Received Date04/10/2018
Supplement Report FDA Received Date04/27/2018
05/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-