WRIGHTS LANE SYNTHES USA PRODUCTS LLC 12MM/130 DEG TI CANN TFNA 400MM/RIGHT-STERILE; ROD,FIXATION,INTRAMEDULLARY
|
Back to Search Results |
|
| Model Number |
04.037.260S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Patient id, date of birth and weight were not provided for reporting.Date of event is unknown.Date of implant is an unknown date in (b)(6) 2017.Device is expected to be returned for manufacturer review/investigation, but has not been received yet.(b)(4).The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2018, during the patient's 4th month post-operative clinic visit, x-rays confirmed that the implanted 12 mm synthes trochanteric fixation nail-advanced (tfna) broke at the juncture of the nail and the lag screw.The patient was originally implanted with the device on an unknown date in (b)(6) 2017.On (b)(6) 2018, the device was explanted due to non-union and delayed healing, and the patient was revised to a 130 angled blade plate.Fragments generated from the broken device were easily removed.Surgery was completed successfully with no surgical delay and no further medical or surgical intervention was required.No other information available.Concomitant devices reported: tfna screw 105 mm (part # 04.038.105, lot # 7939604, qty: 1).Unknown synthes tfna lag screw (part # unknown, lot # unknown, qty: 1).Unknown synthes 5.0 mm distal locking screw (part # unknown, lot # unknown, qty: 1).This report is for one (1) 12 mm/130 deg ti cann tfna 400 mm/right-sterile.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Subject device has been received and is currently in the evaluation process.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device history records review was completed for part# 04.037.260s, lot# h247145.Manufacturing location: monument, manufacturing date: dec 07, 2016, expiry date: oct 31, 2026.Component parts reviewed: 04.037.942.2 - lock prong 130 degree, tfna bp-55 lot - l103742, 04.037.912.4 - wave spring, shim ended bp-55 lot ¿ 9937550, 04.037.912.3 - tfna lock drive bp-58 lot - h210068, 21127 - raw material lot bp-80 lot - h123944.Raw material received from metalwerks pmd, inc.Certificate of analysis for titanium received from metalwerks met specification.Raw material receiving/putaway checklist met specification.Inspection sheet for in-process/inspect dimensional/final and inspection sheet - tfna assembly inspection met inspection acceptance criteria.No non- conformance reports were generated during production.Review of the device history records showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Sterility documentation was reviewed and determined to be conforming.Product development investigation was completed.The returned nail (04.037.260s, h247145) is part of the tfna system for intramedullary fixation of proximal femoral fractures.The returned nail was received broken at the region through which head elements are intended to be inserted, which confirmed the complaint condition and made further replication inapplicable.Relevant drawings were reviewed and determined to be suitable for the intended design, application, and dimensional conformity when used and handled as recommended.No material review report, non- conformance reports, or actions related to the complaint condition were generated during production of the returned part.Review of device history record showed that there were no issues during the manufacture of the product, which would contribute to the complaint condition.The material properties of the returned part were determined to be conforming at the time of manufacture and based on review of the associated/available dhr and based on the details of the complaint and investigation of the returned part, additional material testing is not required.During the investigation, no unidentified product design issues or discrepancies were observed that may have contributed to the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Dimensional check (ca215p) for proximal diameter measured ø15.68-15.69 and found confirming with specification of ø15.66 ± 0.1mm.Based on the available information it is not possible to determine a definitive root cause for the complaint condition.Some potential root causes include patient compliance, bone health, or other potential comorbidities which could have made the patient susceptible to instability and poor bone healing.Also, if the selected implants were less than ideal for the patient¿s anatomy and or not implanted using the intended procedure, it could have contributed to the complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|