• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. EVERA XT VR; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. EVERA XT VR; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER Back to Search Results
Model Number DVBB1D1
Medical Device Problem Codes Inappropriate/Inadequate Shock/Stimulation (1574); Invalid Sensing (2293)
Health Effect - Clinical Codes Atrial Fibrillation (1729); Dizziness (2194); Shock from Patient Lead(s) (3162)
Date of Event 02/09/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant products: 0085 lead, implanted: (b)(6) 2004.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient was inappropriately shocked due to atrial driven arrhythmias detected as ventricular fibrillation (vf) by the implantable cardioverter defibrillator (icd).The patient¿s family member called to report an inappropriate shock with the patient ending up in the emergency room (er).The patient¿s family member reports the patient having atrial fibrillation (af) and symptoms of dizziness.A follow up with the patient¿s healthcare provider yielded that the icd was reprogrammed and the issue was resolved.The icd remains in use.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERA XT VR
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key7413888
Report Number3004209178-2018-06978
Device Sequence Number962985
Product Code LWS
UDI-Device Identifier00643169530300
UDI-Public00643169530300
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P980016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/14/2018
Device Model NumberDVBB1D1
Device Catalogue NumberDVBB1D1
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2018
Initial Report FDA Received Date04/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
6935-65 LEAD
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age31 YR
-
-