| Model Number |
694775 |
| Medical Device Problem Codes |
Failure to Capture (1081); Fracture (1260); High impedance (1291); Over-Sensing (1438); High Capture Threshold (3266)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/16/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated oversensing due to non-physiologic signals/sensing integrity counter.Analysis of the device memory indicated oversensing associated with the right ventricular lead.Analysis of the device memory indicated the criteria for right ventricular lead integrity alert were met.Analysis of the device memory indicated the impedance of the right ventricular pacing lead was beyond the expected upper range.Analysis of the device memory indicated an alert for lead impedance out of range.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that right ventricular (rv) lead exhibited high thresholds, loss of capture, high and undefined pacing impedances and oversensing.A possible lead fracture was suspected.It was noted that a lead integrity alert was triggered, a rv bipolar lead impedance alert was triggered, and that there was a sensing issue due to number of short v-v intervals.The rv lead was capped and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was further reported that this capped right ventricular (rv) lead had been partially explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the proximal portion of the lead was returned, analyzed.Analysis indicated the distal conductor of the lead developed a fracture due to flexing while in vivo.The interior superior vena cava defibrillation cable developed a fracture at the first rib/clavicle location while in vivo.The proximal conductor of the lead developed a fracture due to flexing while in vivo.The right ventricular defibrillation coil was fractured at the 1st rib/clavicle location.The inner insulation of the lead developed a breach at the first rib/clavicle location while in vivo.The outer insulation of the lead developed a breach at the first rib/clavicle location while in vivo.The outer insulation of the lead was breached due to bi/multi-lumen tubing voids while in vivo.The overlay tubing was breached at the 1st rib/clavicle location.If information is provided in the future, a supplemental report will be issued.
|
| |
|