Model Number 6935M72
Medical Device Problem Codes Signal Artifact/Noise (1036); Break (1069); Fracture (1260); Over-Sensing (1438)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/01/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated oversensing associated with the right ventricular lead.Analysis of the device memory indicated the criteria for the right ventricular lead integrity alert were met.Analysis of the device memory indicated the impedance of the right ventricular pacing lead was beyond the expected upper range.Analysis of the device memory indicated oversensing due to non-physiologic signals/sensing integrity counter.Concomitant medical products: fr995-29 tissue valve, implanted: (b)(6) 2006.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that a lead integrity alert (lia) triggered for the right ventricular (rv) lead.The lead exhibited noise, elevated sensing integrity counter (sic), oversensing, and was possibly fractured.The lead was turned off and later explanted and replaced.During the lead replacement procedure, the physician noticed blood and fluid underneath the insulation of the lead.The physician was unable to locate where the insulation breach occurred.The lead was explanted and replaced.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE S
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key7414912
Report Number2649622-2018-03669
Device Sequence Number1446371
Product Code LWS
UDI-Device Identifier00643169309784
UDI-Public00643169309784
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/14/2017
Device Model Number6935M72
Device Catalogue Number6935M72
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/06/2018
Initial Report FDA Received Date04/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DVMB1D4 ICD,
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age60 YR
-
-