| Model Number |
6935M72 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Break (1069); Fracture (1260); Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/01/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated oversensing associated with the right ventricular lead.Analysis of the device memory indicated the criteria for the right ventricular lead integrity alert were met.Analysis of the device memory indicated the impedance of the right ventricular pacing lead was beyond the expected upper range.Analysis of the device memory indicated oversensing due to non-physiologic signals/sensing integrity counter.Concomitant medical products: fr995-29 tissue valve, implanted: (b)(6) 2006.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that a lead integrity alert (lia) triggered for the right ventricular (rv) lead.The lead exhibited noise, elevated sensing integrity counter (sic), oversensing, and was possibly fractured.The lead was turned off and later explanted and replaced.During the lead replacement procedure, the physician noticed blood and fluid underneath the insulation of the lead.The physician was unable to locate where the insulation breach occurred.The lead was explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|