| Model Number |
D334TRG |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/26/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Continuation of concomitant products: 429688 lead, implanted: (b)(6) 2015.Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was evaluated in clinic and the cardiac resynchronization therapy defibrillator (crt-d) had reached premature elective replacement indicator (eri).The left ventricular (lv) lead also showed a high pacing threshold, along with inconsistent capture.The lv output was increased and the device was later explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|