| Model Number |
D204TRM |
| Medical Device Problem Codes |
Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
02/15/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the implantable cardioverter defibrillator (icd) inappropriately shocked the patient due to oversensing of noise.The patient was undergoing surgery and electrocautery was used without applying a magnet to the device.The device detected the electrocautery as noise.The device remains in use.No further patient complications have been reported as a result of this event.
|
| |
|