| Catalog Number |
00025000115 |
| Medical Device Problem Code |
Device Handling Problem (3265)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
03/16/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A customer reported that their finger was cut when they snapped open a vial of gem cvp1 with co-ox.The cut was minor; the wound was washed and dressed.An instrumentation laboratory (il) representative contacted the facility.The facility verified that protocol was not followed; the operator should have used the ampoule breaker.Instrumentation laboratory's review of the operator's manual and instruction for use sheet confirmed appropriate instruction for the use of the ampoule breaker to snap open the ampoule.No remedial action is required.
|
| |
|
Event or Problem Description
|
|
It was reported that a customer cut their finger while snapping open a cvp vial.The customer reported that it was a minor cut.He removed gloves to wash hands with soap and water, then used a bandage.
|
| |
|