• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND ATTAIN COMMAND; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND ATTAIN COMMAND; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 6250V-MP
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product event summary : the partial delivery catheter was returned and analyzed.Analysis indicated that the mechanical operation of the catheter peeling/slitting/splitting showed a spiral slit.Visual analysis of the lead indicated damage during use.The analyst noted that the catheter was returned without the dilator.Visual inspection found there was severe damaged.It appeared an improper slitting on the hub, and there was several spiral slitting on the shaft, damage during the procedure.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during an implant procedure of the left ventricular lead, the hub of the catheter could not be slit.Two slitters were attempted and then scissors were used.It was noted the lead did not dislodge.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN COMMAND
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7415749
Report Number9612164-2018-00716
Device Sequence Number1482385
Product Code DQY
UDI-Device Identifier00643169142893
UDI-Public00643169142893
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K123153
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/11/2019
Device Model Number6250V-MP
Device Catalogue Number6250V-MP
Device Lot Number0008916663
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2018
Initial Report FDA Received Date04/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age78 YR
Patient Weight129
-
-