| Model Number |
DTBB1QQ |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/08/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 407652 lead, implanted on (b)(6) 2005.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
In response to the advisory noting the potential for internal device arcing during high-voltage charging, the cardiac resynchronization therapy defibrillator (crt-d) was prophylactically replaced.No device malfunction was observed while the device was in use, however, given the potential for loss of device functionality, the crt-d was explanted and replaced to preclude harm to the patient.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
This device was reported as included in the field action noted and returned product investigation found the device performed as described in the field action.Evaluation summary: the device was returned and analyzed.Device passed initial functional test, however the device failed extended high voltage cycle testing and failed residual gas analysis for internal device minimum nitrogen content.If information is provided in the future, a supplemental report will be issued.
|
| |
|