| Model Number |
D314VRG |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Failure to Interrogate (1332)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/21/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the implantable cardioverter defibrillator (icd) exhibited premature battery depletion and was unable to be interrogated.The device was explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was returned and analyzed.The returned device indicated shorting/low impedance in the battery.Hybrid analysis confirmed a severely depleted battery.When powered by an external supply, the system current drain was nominal, and the device was fully functional throughout the analysis.No battery depletion mechanism, such as high system current drain, was observed.No hybrid anomalies were found.Battery analysis revealed that lithium-plating breaches were found along the top edge of the anode separator in segment 2, adjacent to the exposed cathode tab.Plated-lithium extended from the breached opening and touched the cathode tab.Based on this analysis, the premature battery depletion resulted from an internal lithium-plating short.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|