MAUDE Adverse Event Report: COOK INC TURBO-JECT SINGLE LUMEN PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER SET; LJS TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC TURBO-JECT SINGLE LUMEN PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER SET; LJS TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER Back to Search Results
Model Number G55883
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: the date of event is estimated to be (b)(6) 2018.(b)(4).Device evaluated by manufacturer - to date, it is unknown if the device will be returned for evaluation.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported a leak was observed at the purple hub on a picc line when medication was administered at home.The patient was admitted to the emergency room and a replacement line was placed the following day.As reported, the picc line was used to deliver total parental nutrition (tpn) and a tiny hole was found during the replacement procedure.
 
Additional Manufacturer Narrative
Investigation - evaluation: a review of the complaint history, device history record, documentation, instructions for use (ifu), manufacturing instructions, quality control, and specifications of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.There is no indication that a design or process related failure mode contributed to this event.Current controls for manufacturing are in place to assure functionality and device integrity prior to shipping.Review of production and quality documentation did not observe any specific issues with current manufacturing or quality controls that may have contributed to this incident.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.
 
Event or Problem Description
No new event description information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURBO-JECT SINGLE LUMEN PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER SET
Common Device Name
LJS TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7416736
Report Number1820334-2018-00826
Device Sequence Number11003608
Product Code LJS
UDI-Device Identifier00827002558830
UDI-Public(01)00827002558830(17)190816(10)8147139
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K111244
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberG55883
Device Catalogue NumberUPICS-3.0-CT-NT-1110
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/19/2018
Supplement Date Received by Manufacturer06/25/2018
Initial Report FDA Received Date04/10/2018
Supplement Report FDA Received Date07/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age13 YR
-
-