During evaluation of a returned mclean-ring ultrathane enteral feeding tube set, it was noted that the catheter's tip was missing.The tip was not returned; however, as of the date of this report, the investigation and evaluation are ongoing.
|
|
A review of the complaint history, drawings, dimensional verification, device history record, documentation, instructions for use (ifu), manufacturing instructions, specifications, quality control, and visual inspection of the returned device was conducted during the investigation.A document-based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.Review of the device history record of the finished product shows no nonconforming events that could contribute to this failure mode.A complaint history search revealed that there were no other reported complaints for this lot number.Device failure analysis was conducted and identified the tip to be missing.With the missing tip, measuring the proper catheter length is not possible and therefore we cannot confirm whether the device is conforming or not.Further inspection of separation location, it was identified that the edge/face is very smooth and not jagged.This gives evidence that he tip was intentionally cut by the customer prior to procedure.Document review shows many procedures in place that should have identified/prevented the failure mode from being distributed.Therefore the root cause is most likely product use/handling and not manufacturing.Per the [quality engineering] risk assessment, no further action is required.Monitoring will continue to be performed for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.H3 other text : a review of the complaint history, drawings, dimensional verification, device history record, documentation, instructions for use (ifu), manufacturing instructions, specifications, quality control, and visual inspection of the returned device was conducted during the investigation.A document-based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.Review of the device history record of the finished product shows no nonconforming events that could contribute to this failure mode.A complaint history search revealed that there were no other reported complaints for this lot number.Device failure analysis was conducted and identified the tip to be missing.With the missing tip, measuring the proper catheter length is not possible and therefore we cannot confirm whether the device is conforming or not.Further inspection of separation location, it was identified that the edge/face is very smooth and not jagged.This gives evidence that he tip was intentionally cut by the customer prior to procedure.Document review shows many procedures in place that should have identified/prevented the failure mode from being distributed.Therefore the root cause is most likely product use/handling and not manufacturing.Per the [quality engineering] risk assessment, no further action is required.Monitoring will continue to be performed for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|