| Model Number |
DVAB1D1 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the implantable cardioverter defibrillator (icd) was attempted to be implanted but not used.The setscrew would not tighten lead.The icd was removed and a new device was implanted.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned and analyzed.Returned product analysis was performed and no anomalies were found.The returned device indicated a loose/detached set screw.The returned device indicated a set screw with a rounded socket.The returned device indicated the set screw was found in the connector bore.If information is provided in the future, a supplemental report will be issued.
|
| |
|