Model Number 407658
Medical Device Problem Codes Signal Artifact/Noise (1036); Over-Sensing (1438)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during the implant procedure, the lead helix was taken out of the package and exercised prior to being placed in the body.Once placed and before the helix was extended, the signal obtained looked non-physiologic and "noisy".Occasional deflections would appear amidst the noisy baseline and seemed to line up with surface qrs complexes.Recognizing that the helix was still retracted, and that the anatomical location was desirable, the physician extended the helix in the usual fashion to affix it to the cardiac tissue.At this point, the noisy baseline was gone, and clear rv electrograms were noted.Desiring larger r-waves, the physician retracted the helix once again and then placed it on the rv septum where r-waves were larger.The helix was once again deployed for the final time and sensing/impedance/thresholds remained stable throughout the case.The lead was left in the final position and remains in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSUREFIX NOVUS
Common Device Name
ELECTRODE, PACEMAKER, PERMANENT
Manufacturer (Section D)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SG  486056
MDR Report Key7417835
Report Number3008973940-2018-00661
Device Sequence Number1083729
Product Code DTB
UDI-Device Identifier00763000014650
UDI-Public00763000014650
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P930039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/06/2019
Device Model Number407658
Device Catalogue Number407658
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/19/2018
Initial Report FDA Received Date04/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age70 YR
-
-