| Model Number |
6947M97 |
| Medical Device Problem Codes |
Over-Sensing (1438); Decreased Sensitivity (2534); Capturing Problem (2891); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/01/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient¿s right ventricular (rv) lead was experiencing rising threshold, diminished sensing, and t-wave oversensing (twos).A micro-dislodgement of the rv lead was suspected.The rv lead remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was further reported that the rv lead was partially explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the full lead was returned in segments, analyzed.Analysis indicated that the exposed superior vena cava defibrillation coil developed a fracture due to flexing while in vivo.If information is provided in the future, a supplemental report will be issued.
|
| |
|