MAUDE Adverse Event Report: COLOPLAST A/S (2BO)TITAN OTR INFRA ZERO ANG; INFLATABLE PENILE PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S (2BO)TITAN OTR INFRA ZERO ANG; INFLATABLE PENILE PROSTHESIS Back to Search Results
Model Number ESR8222400
Health Effect - Clinical Code No Information (3190)
Date of Event 03/16/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information of conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
 
Event or Problem Description
According to the available information, extrusion.
 
Additional Manufacturer Narrative
This follow-up mdr is created to document the conclusion of the investigation.Quality was unsuccessful in securing the explanted prosthesis for evaluation.Quality reviewed the manufacturing records for this device and confirmed that there were no discrepancies that would have contributed to this complaint and verified that the devices from this lot met all specifications prior to release.Review of nonconformances revealed no nonconformance with this lot that would have contributed to the event.Based on the available information, quality concludes that the extrusion originated from source(s) other than the device.Without the benefit of analyzing the explant, quality cannot confirm any observations and cannot comment on the condition of the prosthesis.If the explanted device becomes available, or additional information is received, quality will re-evaluate this complaint in accordance to procedures.Management routinely reviews events such as this and monitors complaint levels.Additionally, events of this type are captured in the product risk documentation.Based on this information, no further corrective action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
(2BO)TITAN OTR INFRA ZERO ANG
Common Device Name
INFLATABLE PENILE PROSTHESIS
Manufacturer (Section D)
COLOPLAST A/S
holtedam 1
humlebaek, 3050
DA  3050
MDR Report Key7418207
Report Number2125050-2018-00288
Device Sequence Number16579475
Product Code FHW
UDI-Device Identifier05708932487580
UDI-Public05708932487580
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P000006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberESR8222400
Device Catalogue NumberESR8222400
Device Lot Number3670891
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/16/2018
Supplement Date Received by Manufacturer03/16/2018
Initial Report FDA Received Date04/10/2018
Supplement Report FDA Received Date06/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age69 YR
-
-